Nexera Prep - Especificaciones
Preparative Purification Liquid Chromatograph
- Descripción general
- Características
- Aplicaciones
- Especificaciones
- Descargas
- Configuración
- Columnas
- Opciones
- Soporte
Excellent System Expandability
Solvent Delivery Unit Accommodates a Wide Range of Recovery Volumes
Guidelines for Preparative Scale and Maximum Load
The figure at right shows the guidelines for total component capacity with a 250 mm long column when the target component is highly soluble in the mobile phase, separates from impurities, and ions are suppressed. For isocratic elution, in principle, the total component capacity is proportional to the column volume.
LC-40D
• This solvent delivery unit can handle flow rates ranging from those used in analytical scale to those used in semi-preparative (up to 10 mL/min).
• High-precision analysis is possible even in the semi-micro flow-rate range.
Guidelines for Preparative Scale and Maximum Load
Solvent delivery method |
Parallel-type double plunger |
---|---|
Plunger capacity |
10 μL |
Maximum discharge pressure |
44 MPa |
Flow rate setting range |
0.0001 to 5.0000 mL/min (1.0 to 44 MPa) |
Flow rate accuracy |
No more than ±1% or ±2 μL/min, whichever |
Flow rate precision |
No more than 0.06% RSD or 0.02 min SD, whichever is greater |
Constant pressure solvent delivery |
Supported |
Plunger rinsing mechanism |
Optional available |
Operating temperature range |
4 to 35 °C |
Size and weight |
W260 × D500 × H140 mm, 10 kg |
LC-20AR
This solvent delivery unit can handle flow rates used in semi-preparative scale (up to 20 mL/min).
Using a recycle kit enables semi-preparative recycling.
Supports Semi-Preparative and Recycle Preparative
Solvent delivery method |
Parallel-type double plunger |
---|---|
Plunger capacity |
47 μL |
Maximum discharge pressure |
49 Mpa |
Flow rate setting range |
0.001 to 20.000 mL/min |
Flow rate accuracy |
No more than ±1% or ±10 μL/min,whichever is greater (0.1 to 5.0 mL/min) |
Flow rate precision |
No more than 0.08% RSD or 0.02 min SD, whichever is greater |
Constant pressure solvent delivery |
Supported |
Plunger rinsing mechanism |
Syringe or rinsing pump |
Operating temperature range |
4 to 35 °C |
Size and weight |
W 260 x H140 x D 500 mm,16 kg |
LC-20AP
High flow rates (up to 150 mL/min) enable highly efficient, large-scale preparative fractionation.
Large-scale prep solvent delivery fully supports the preparative fractionation workflow including reinjection to assess purity.
Combine with an FCV-200AL low-pressure gradient unit to perform gradient analysis using up to four mobile phases.
Supports Large-Scale Preparative Fractionation
Solvent delivery method |
Parallel-type double plunger |
---|---|
Plunger capacity |
250 μL |
Maximum discharge pressure |
42 Mpa |
Flow rate setting range |
0.01 to 150.00 mL/min |
Flow rate accuracy |
No more than ±1% (1 mL/min, 10 MPa) |
Flow rate precision |
No more than 0.1% RSD or 0.02 min SD,whichever is greater |
Constant pressure solvent delivery |
Supported |
Plunger rinsing mechanism |
Syringe or rinsing pump |
Operating temperature range |
4 to 35 °C |
Size and weight |
W 260 x H210 x D 500 mm,19 kg |