TAC-60L/60i - Especificaciones

Automatic Cooling Attachments

Add Bookmark

TAC-60L

TAC-60L/60i

DSC-60A + TAC-60L + TA-60WS PC set + FC-60A + ATG-60, etc.

The TAC-60L system uses liquid nitrogen as the catalyst for cooling samples down to -130 °C.

- Stable Baseline
The tank internal pressure is optimally controlled according the residual amount of liquid nitrogen in the tank to keep the supply flowrate constant. Pulsation is reduced, and the DSC baseline is stable even during cooling.

- Space-Saving
Installation space is saved since the liquid nitrogen supply control system is placed compactly in the pump head section.

- Refrigerant Savings
Samples can be cooled only when required since supply start/stop is controlled in steps according to a temperature program. Also, for safe operation and minimizing refrigerant consumption, refrigerant supply is automatically stopped when the measuring unit reaches a high temperature.

- Autosampler Continuous Measurement Supported
Samples can be replaced easily during cooling thanks to the unique layout at the rear of the cooling unit. Moreover, gas can be allowed to pass for a fixed period of time after measurement ends to prevent dew and frost condensation. (when the optional ATG-60 is connected)

TAC-60L Liquid Nitrogen Cooling System

Temperature range

-130 °C to 500 °C

Cooling capacity

-10 °C/min (to -80 °C)

Cooling program

Start/stop of liquid nitrogen supply can be set to operate in sync with the temperature program.

Protective functions

Gas flow start/stop after cooling stops (ATG-60), liquid nitrogen supply stop at high temperatures

* Including accessory liquid nitrogen tank (35 L). The continuous operating time when this tank is used is approximately 8 hours. (This changes according to the temperature program.)

Leer más
Ver menos

TAC-60i

TAC-60L/60i

DSC-60A + TAC-60i + TA-60WS PC set + FC-40A + cooling unit

The TAC-60i electric-cooling type attachment does not require liquid nitrogen and can cool samples to down -50 °C.

- No Catalyst Required
Liquid nitrogen or other catalysts are not used, which means that liquid nitrogen need not be prepared before measurement or that there is no worry about ventilation.

- Just Connect the Cooling Unit
Samples can be cooled down to -50 °C for measurement by simply connecting a cooling unit (intra cooler).

TAC-60i Electric-Cooling Attachment

Temperature range

-50 °C to 500 °C

Cooling capacity

-10 °C/min
(at -10 °C)
-5 °C/min
(at -30 °C)

Protective functions

DSC heating furnace forcibly turned OFF at a cooling head temperature of 70 °C

*To prevent condensation, be sure to use dry nitrogen or inert gas when operating this system.

{"title":"Descargas de la hoja de especificaciones","description":"Descargue las \u00faltimas hojas de especificaciones.","source":"product","key":3512,"max":"30","filter_types":["spec_sheet"],"link_title":"Ver otras descargas","link_url":""} {"title":"Descargas","description":"Descargue el folleto m\u00e1s reciente.","source":"product","key":3512,"max":"30","filter_types":["brochures"],"link_title":"Ver otras descargas","link_url":"","pdf_links":[]}