SALD-2300 - Especificaciones
Laser Diffraction Particle Size Analyzer
- Descripción general
- Características
- Aplicaciones
- Especificaciones
- Descargas
- Configuración
- Opciones
- Soporte
SALD-2300
General Specifications
Measurement principle |
Laser Diffraction Method |
---|---|
Measurement range |
17nm(0.017μm) to 2500μm |
Measurement range |
17nm(0.017μm) to 400μm |
Measurement range |
30nm(0.030μm) to 280μm |
Measurement range |
300nm(0.3μm) to 2500μm |
Note 1: The measurement range varies according to the shape etc. of the particle.
SALD-2300
Measurement Unit: SALD-2300 (P/N: 347-61700-42[115V], 347-61700-44[230V])
Light source |
Red Semiconductor laser (Wavelength 680 nm) |
---|---|
Light detector |
Detector elements for UV semiconductor laser |
System Compliance |
Class 1 Laser Product, CE compliant |
Required power supply |
115 or 230 VAC as ordered 100 VA |
Dimensions & weight |
W680mm×D280mm×H430mm, 31kg |
Operating Environment |
Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Note 2: Reference sample and USB cable (2 m) supplied as standard
SALD-MS23
Sampler: SALD-MS23 (P/N: 347-61701-42[115V], 347-61701-44[230V])
Dispersing bath |
Capacity: 100~280cm3 |
---|---|
Sonicator |
Frequency about 32 kHz, output about 40 W |
Liquid Pump |
Radial pump, maximum flow rate 2000cm3/min |
Liquid Pump Material |
Stainless (SUS 304, SUS 316), Tetrafluoroethylene (PTFE), Perfluoroelastmor (FEP) or Kalrez, Thermoflon Pascal (inside) |
Liquid Supply Pump |
Diaphragm pump, maximum flow rate 750cm3/min |
Liquid Supply Pump Material |
Tetrafluoroethylene, polyvinyl dene fluoride |
Flow Cell |
Quartz glass |
Required power supply |
115 or 230 VAC as ordered, 200 VA |
Dimensions & weight |
W390mm×D520mm×H430mm,18kg |
Operating Environment |
Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Note 3: USB cable (2 m) supplied as standard
SALD-DS5
Cyclone Injection Type Dry Measurement Unit: SALD-DS5 (P/N: 347-61706-42[115V], 347-61706-44[230V])
System |
Cyclone type |
---|---|
Communication method |
USB (PC control) |
Required power supply |
115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor) |
Dimensions & weight |
W 240×D 310×H 210 mm, 10 kg |
Operating Environment |
Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
SALD-DS5
Cyclone Injection Type Dry Measurement Unit: SALD-DS5 (P/N: 347-61706-42[115V], 347-61706-44[230V])
Primary pressure |
0.6 to 0.8 MPa |
---|---|
Secondary pressure |
0.05 to 0.5 MPa |
Filtration rating |
Removing particles of 5 μm or larger |
Connection to air source |
Tube of 6 mm outside diameter |
Communication method |
USB (PC control) |
Required power supply |
115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor) |
Dimensions & weight |
W 130×D 223×H 233 mm, 3 kg |
Operating Environment |
Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Compressor: Output |
0.4 kW |
---|---|
Compressor: Minimum pressure |
Approx. 7 kgf/cm2 (approx. 0.69 MPa) |
Compressor: Air discharge |
45 L/min |
Compressor: Tank volume |
Approx. 30 L |
Dust collector: Type |
Vacuum cleaner (Paper pack type) |
Dust collector: Dust collecting efficiency |
More than 99% for 0.3 μm particles |
Dust collector: Capacity |
2.0 m3/min or more |
Dust collector: Vacuum |
Approx. 2000 mm Aq or lower |
Dust collector: Suction hose diameter |
Approx. 32 mm |
WingSALD II
Measurement and Data Display Functions
Measurement of Particle Size Distribution |
Allows measurements using measurement assistant function (interactive process based on SOP) |
---|---|
Refractive Index Setting |
Automatic refractive index calculation function (LDR method: Light Intensity Distribution Reproduction Method) makes setting the refractive index easy. |
Real-Time Display |
Particle size distribution/light intensity distribution simultaneous display |
Diagnostics/Adjustments |
Self-diagnostic function and cell check function |
Recalculation of Particle Size Distribution |
Batch recalculation of max. 200 distributions |
Display of Particle Size Distribution Data |
Displays overlay of max. 200 distributions |
Display of Light Intensity Distribution |
Displays overlay of max. 200 distributions |
Statistical Data Processing |
Max. 200 sets of data (also allows overlaying max. 200 data sets) |
Time-Series Processing |
Max. 200 sets of data |
3-Dimensional Graphing |
Max. 200 sets of data |
Data Transfer via Clipboard |
[Image Output]: Outputs entire data sheet or graph only. |
Data Sorting |
Sorts by file name, sample ID, sample number, or refractive index |
WingSALD II
Output Conditions
Particle Size (μm) Divisions |
Fixed 51 or 101 divisions : User settable 51 divisions |
---|---|
Particle Amount (%) Divisions |
Fixed 51 divisions : User settable 51 divisions |
Distribution Basis |
Count, length, area, or volume |
Expression of Cumulative Distribution |
Oversized or undersized |
Expression of Frequency Distribution |
q, q / Δ×, q / Δlog × |
Smoothing Levels |
10 levels |
Distribution Function Fitting |
Rosin-Rammler distribution, logarithmic Gaussian distribution |
Data Shifting |
±10 levels |
Report Function |
Single data sets, overlaid data, statistical data, time-series data, or 3D data can be selected and output using batch processing |
WingSALD II
Data Analysis Functions
Scattering Angle Evaluation Function |
Evaluates scattering characteristics within micro angle regions for samples such as optical films and sheets. |
---|---|
Data Emulation Functions |
Emulates measurement results from other instruments and measurement principles, using SALD series measurement results. |
Mixture Data Simulation Function |
Simulates particle size distributions using any mixture ratio of multiple particle size distributions. |
Data Connection Function |
Combines two particle size distributions with different measurement ranges at any particle size point to create a single particle size distribution. |
Continuous Measurement Function |
Continuously measures changes in particle size distributions and particle diameters over time, at intervals as short as one second, and saves the results. |
Note 4:The LDR method automatically calculates an appropriate refractive index based on consistency between the actual measured light intensity distribution and one reproduced (recalculated) from particle size distribution data. This method was developed by Shimadzu and published in two technical papers. It is sometimes called the "Kinoshita Method", in academic communities, after the name of Shimadzu’s engineer.
SALD-2300
Measurement Unit: SALD-2300 (P/N: 347-61700-42[115V], 347-61700-44[230V])
Measurement range |
17 nm to 2,500 µm |
---|---|
Light source |
Red semiconductor laser |
Detection elements |
84 elements |
Options |
Multifunctional variable-volume sampler, batch cell, highconcentration sample measurement system, cyclone injection type dry measurement unit |
- Descripción general
- Características
- Aplicaciones
- Especificaciones
- Descargas
- Configuración
- Opciones
- Soporte