SSG Series Strain Gauge Extensometers - Especificaciones
SSG10-10
Measurement range (mm/FS) |
1.0,0.5,0.2,0.1 |
---|---|
Gauge length (mm) |
10 |
Max. measurement range (mm) |
1 |
Max. specimen size (mm) : Rod maximum diameter |
φ10 |
Max. specimen size (mm) : Flat plate width x thickness |
W15×T10 |
Accuracy |
JIS Class 1 ASTM Class CMeasurement range up to 0.3 mm: ± 3 μmMeasurement range above 0.3 mm to max.: ±1.0% of indicated value |
Operational temp. range |
5 to 40°C (JIS guaranteed precision range: Fluctuation within ±2°C at 18 to 28°C) |
Power Requirements |
ESA type amplifier: Single-phase 100 V to 240 V AC at 150 VA |
SSG25-10
Measurement range (mm/FS) |
2.5,1.25,0.5,0.25 |
---|---|
Gauge length (mm) |
25 |
Max. measurement range (mm) |
2.5 |
Max. specimen size (mm) : Rod maximum diameter |
φ10 |
Max. specimen size (mm) : Flat plate width x thickness |
W15×T10 |
Accuracy |
JIS Class 1 ASTM Class CMeasurement range up to 0.3 mm: ± 3 μmMeasurement range above 0.3 mm to max.: ±1.0% of indicated value |
Operational temp. range |
5 to 40°C (JIS guaranteed precision range: Fluctuation within ±2°C at 18 to 28°C) |
Power Requirements |
ESA type amplifier: Single-phase 100 V to 240 V AC at 150 VA |
SSG50-10
Measurement range (mm/FS) |
5,2.5,1.0,0.5 |
---|---|
Gauge length (mm) |
50 |
Max. measurement range (mm) |
5 |
Max. specimen size (mm) : Rod maximum diameter |
φ15 |
Max. specimen size (mm) : Flat plate width x thickness |
W25×T14 |
Accuracy |
Measurement range 0 to 0.13 mm: ±1 μmMeasurement range 0.13 mm to 0.3 mm: ±3 μmMeasurement range 0.3 mm to 5 mm: ±1.0% of indicated value |
Operational temp. range |
5 to 40°C (JIS guaranteed precision range: Fluctuation within ±2°C at 18 to 28°C) |
Weight |
about 75g |
Power Requirements |