0

    4890 Controller - Especificaciones

    Servopulser Control Unit

    Agregar marcador

    4890 Controller

    Logical operation resolution

    32 bits

    Loading waveform

    Sine, haversine, triangular, hold rectangular, and combinations of these

    Special waveforms

    Slow start/stop, sweep, arbitrary

    Frequency

    0.00001 to 500 Hz

    Ramp speed

    1×10-9 to 100 full scale/sec

    Hold time

    0.01 to 1×109 sec

    Cycle counter

    1010

    Measurement amplifier type

    Test force/displacement (standard accessory)/strain/elongation/voltage inputs
    * 6 additional optional amplifiers can be added.

    Measurement amplifier functions

    Rangeless, auto-zero function, automatic recognition and calibration functions

    Measured value display

    Instantaneous values or upper/lower limit peaks values are displayed in engineering units

    Number of control channels

    Max. 8

    Control functions

    2-degree-of-freedom PID control
    Auto-tuning function
    Realtime gain adjustment function
    Logical operation control functions (virtual transducer)
    Automatic amplitude/mean-value correction function (AGC)
    Shockless control switching function
    Function to prevent overloading during specimen clamping (sub controller function)

    Alarm functions

    Excessive amplitude detection, low amplitude detection, breakage sensitivity, breakage detection
    hydraulic unit alarm, preset count, external alarm I/O

    Software

    GLUON software package for Windows

    Size (W × D × H)

    205 × 483 × 466 mm (excluding protrusions)

    Power Requirements

    100 ±10 V 50 Hz/60 Hz, 1.0 kVA single-phase

    Leer más