LC-2030C NT - Especificaciones
HPLC System
Pump
Degassing unit |
Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 µL) |
---|---|
Pumping method |
Parallel-type double plunger |
Pulsation |
≤ 0.1 MPa (1.0 mL/min,10 MPa, Water) |
Flow rate setting range |
0.0001 to 10 mL/min |
Flow rate accuracy |
≤ ±1% or ±2 µL/min, below whichever is greater (0.01 to 2 mL/min, Specified conditions) |
Flow rate precision |
≤ 0.06 %RSD or ≤ 0.02 minSD, below whichever is greater |
Configuration |
Four-solvent low-pressure gradient |
Gradient / range of set concentrations |
0 to 100%, in 0.1% steps |
Gradient / concentration accuracy |
±0.5% (0.1 to 2 mL/min, 1 to 20 MPa, Specified conditions) |
Gradient / concentration precision |
±0.1% (1 mL/min, 10 MPa, Specified conditions) |
Maximum pressure |
Packed column side : 34 MPa (0.0001 to 5 mL/min) , 22 MPa (5.0001 to 10 mL/min) |
System Delay Volume |
approximately 615 µL (Packed column side) |
Autosampler
Injection method |
Total-volume sample injection |
---|---|
Injection volume accuracy |
±1% (50 µL, N=10) |
Injection volume setting range |
0.1 to 100 µL (Option:0.1 to 50 µL) |
Injection volume reproducibility |
RSD ≤0.20% (5.0 to 100 µL), RSD ≤0.25% (2.0 to 4.9 µL) |
Cross-contamination |
0.0025% (Caffeine, Specified conditions) |
Injection cycle time |
Min. 14 sec (Specified conditions) |
Samples for processing |
336 (1 mL), 216 (1.5 mL), 112 (4 mL), 4 (MTP/DWP) |
Sample cooler |
4 to 45°C |
Injection linearity |
≥0.9999 (1 to 100 µL, Specified conditions) |
Column Oven
Heating and cooling method |
Forced air circulation method |
---|---|
Containable column size |
Slide-in column |
Temperature control range |
Room temperature −5 to 60°C, |
Temperature control precision |
±0.1°C |
Temperature stability |
±0.8°C (Specified conditions) |
Flow rate switching valve |
Not available |
PDA Detector
Wavelength range |
190 to 800 nm |
---|---|
Spectral resolution |
1.4 nm (Specified condition) |
Slit width |
1.2 nm, 8 nm |
Device resolution |
0.6 nm/pixel |
Number of photodiode array elements |
1024 |
Wavelength accuracy |
≤ ±1 nm |
Noise level |
≤ ±3 × 10−6 AU (250 nm, reference: 350 nm, Specified conditions) |
Drift |
≤ 500 × 10−6 AU/h (250 nm, reference: 350 nm, Specified conditions) |
Linearity |
Up to 2 AU (5%) |
Sampling rate |
Up to 100 Hz |
Light source |
Deuterium (D2) lamp (Standard), tungsten (W) lamp (option) |
Flow Cell |
8 µL (10mm, TC), 12 MPa |
Option cell |
Conventional: 10 µL (10 mm, TC), |
Miscellaneous
Dimensions |
W410×H605×D500 mm |
---|---|
Weight |
67 kg |
Available pH range |
1 to 13 |
Materials for parts in contact with liquids |
Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby, |
Workstation |
LabSolutions™ LC/GC Ver. 5.92 or later, |