AP Series - Especificaciones
Analytical Balances
AP225W
External Calibration Weight Range for Span Calibration |
95 to 220.00090 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg |
Capacity |
220g |
Repeatability (Standard deviation) |
0.015mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 213(W) x 411(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
8 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.1 mg |
Minimum Weight *2 |
20 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
*1 Minimum display 0.01mg models provide two internal weights as standard.
*2 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance’s capacity of 5% (or 5 grams’ weight). In the case of the AP225W, the results are for tests carried out with the optional internal windbreak plate applied. The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.
*3 The response time for displaying 90% of the added sample amount value in trace measurements (from 1mg)
*4 The response time value is typical.
*5 Non-condensing.
AP224W-AD
Linearity *2 |
±0.2 mg |
---|---|
Minimum Weight *2 |
200 mg |
Repeatability (Standard deviation) |
0.1 mg |
External Calibration Weight Range for Span Calibration |
95 to 220.009 g |
Calibration Weight |
Built-in *1 |
Minimum display |
0.1 mg |
Capacity |
220 g |
Response Time for race Measurements *3 |
2 sec. |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Ionizer |
Included |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Display |
OEL display (dot matrix) |
Pan Size |
ø91 mm |
Body Dimensions |
Approx. 215(W) x 367(D) x 346(H)mm |
Weight |
Approx. 7.0 kg |
Touchless Sensors |
Included |
Adjustable Internal Windbreak Plate |
Not Available |
Sample (Concentration) Preparation |
Included |
Automatic Doors |
Included |
Clock-CAL |
Included |
Inspection Support Function |
Included |
mol Conversion Function |
Included |
HPLC Buffer Solution Preparation |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP225WD-AD
Linearity *2 |
±0.2 mg /±0.1 mg |
---|---|
Minimum Weight *2 |
30 mg |
Repeatability (Standard deviation) |
0.1 mg / 0.05 mg |
External Calibration Weight Range for Span Calibration |
95 to 220.009 g |
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg / 0.1 mg |
Capacity |
220 g / 102g |
Response Time for race Measurements *3 |
2 sec. |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Ionizer |
Included |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Display |
OEL display (dot matrix) |
Pan Size |
ø91 mm |
Body Dimensions |
Approx. 215(W) x 411(D) x 346(H)mm |
Weight |
Approx. 7.9 kg |
Touchless Sensors |
Included |
Adjustable Internal Windbreak Plate |
Included |
Sample (Concentration) Preparation |
Included |
Automatic Doors |
Included |
Clock-CAL |
Included |
Inspection Support Function |
Included |
mol Conversion Function |
Included |
HPLC Buffer Solution Preparation |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. / 8 sec |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP125WD-AD
Linearity *2 |
±0.2 mg /±0.05 mg |
---|---|
Minimum Weight *2 |
30 mg |
Repeatability (Standard deviation) |
0.1 mg / 0.02 mg |
External Calibration Weight Range for Span Calibration |
95 to 120.0090 g |
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg / 0.1 mg |
Capacity |
120 g / 52g |
Response Time for race Measurements *3 |
2 sec. |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Ionizer |
Included |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Display |
OEL display (dot matrix) |
Pan Size |
ø91 mm |
Body Dimensions |
Approx. 215(W) x 411(D) x 346(H)mm |
Weight |
Approx. 7.9 kg |
Touchless Sensors |
Included |
Adjustable Internal Windbreak Plate |
Included |
Sample (Concentration) Preparation |
Included |
Automatic Doors |
Included |
Clock-CAL |
Included |
Inspection Support Function |
Included |
mol Conversion Function |
Included |
HPLC Buffer Solution Preparation |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. / 8 sec |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP135W-AD
Linearity *2 |
±0.1 mg |
---|---|
Minimum Weight *2 |
30 mg |
Repeatability (Standard deviation) |
0.05 mg |
External Calibration Weight Range for Span Calibration |
45 to 135.009 g |
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg |
Capacity |
135 g |
Response Time for race Measurements *3 |
2 sec. |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Ionizer |
Included |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Display |
OEL display (dot matrix) |
Pan Size |
ø91 mm |
Body Dimensions |
Approx. 215(W) x 411(D) x 346(H)mm |
Weight |
Approx. 7.9 kg |
Touchless Sensors |
Included |
Adjustable Internal Windbreak Plate |
Included |
Sample (Concentration) Preparation |
Included |
Automatic Doors |
Included |
Clock-CAL |
Included |
Inspection Support Function |
Included |
mol Conversion Function |
Included |
HPLC Buffer Solution Preparation |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
8 sec. |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP225W-AD
Linearity *2 |
±0.1 mg |
---|---|
Minimum Weight *2 |
20 mg |
Repeatability (Standard deviation) |
0.015mg |
External Calibration Weight Range for Span Calibration |
95 to 220.00090 g |
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg |
Capacity |
220g |
Response Time for race Measurements *3 |
2 sec. |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Ionizer |
Included |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Display |
OEL display (dot matrix) |
Pan Size |
ø91 mm |
Body Dimensions |
Approx. 215(W) x 411(D) x 346(H)mm |
Weight |
Approx. 7.9 kg |
Touchless Sensors |
Included |
Adjustable Internal Windbreak Plate |
Included |
Sample (Concentration) Preparation |
Included |
Automatic Doors |
Included |
Clock-CAL |
Included |
Inspection Support Function |
Included |
mol Conversion Function |
Included |
HPLC Buffer Solution Preparation |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
8 sec. |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
*1 Minimum display 0.01mg models provide two internal weights as standard.
*2 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance’s capacity of 5% (or 5 grams’ weight). In the case of the AP225W, the results are for tests carried out with the optional internal windbreak plate applied. The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.
*3 The response time for displaying 90% of the added sample amount value in trace measurements (from 1mg)
*4 The response time value is typical.
*5 Non-condensing.
AP135W
External Calibration Weight Range for Span Calibration |
45 to 135.009 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg |
Capacity |
135 g |
Repeatability (Standard deviation) |
0.05 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 212(W) x 411(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
8 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.1 mg |
Minimum Weight *2 |
30 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP125WD
External Calibration Weight Range for Span Calibration |
95 to 120.0090 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg / 0.1 mg |
Capacity |
120 g / 52g |
Repeatability (Standard deviation) |
0.1 mg / 0.02 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 213(W) x 411(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. / 8 sec |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg /±0.05 mg |
Minimum Weight *2 |
30 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP324Y
External Calibration Weight Range for Span Calibration |
95 to 320.009 g |
---|---|
Calibration Weight |
No |
Minimum display |
0.1 mg |
Capacity |
320 g |
Repeatability (Standard deviation) |
0.15 mg |
Inspection Support Function |
Not Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 6.5 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Not Included |
Clock-CAL |
Not Included |
HPLC Buffer Solution Preparation |
Not Included |
Sample (Concentration) Preparation |
Not Included |
mol Conversion Function |
Not Included |
Recipe Compounding |
Not Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Not Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.3 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) |
AP224Y
External Calibration Weight Range for Span Calibration |
95 to 220.009 g |
---|---|
Calibration Weight |
No |
Minimum display |
0.1 mg |
Capacity |
220 g |
Repeatability (Standard deviation) |
0.1 mg |
Inspection Support Function |
Not Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 6.5 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Not Included |
Clock-CAL |
Not Included |
HPLC Buffer Solution Preparation |
Not Included |
Sample (Concentration) Preparation |
Not Included |
mol Conversion Function |
Not Included |
Recipe Compounding |
Not Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Not Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) |
AP124Y
External Calibration Weight Range for Span Calibration |
45 to 120.009 g |
---|---|
Calibration Weight |
No |
Minimum display |
0.1 mg |
Capacity |
120 g |
Repeatability (Standard deviation) |
0.1 mg |
Inspection Support Function |
Not Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 6.5 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Not Included |
Clock-CAL |
Not Included |
HPLC Buffer Solution Preparation |
Not Included |
Sample (Concentration) Preparation |
Not Included |
mol Conversion Function |
Not Included |
Recipe Compounding |
Not Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Not Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) |
AP324X
External Calibration Weight Range for Span Calibration |
95 to 320.009 g |
---|---|
Calibration Weight |
Built-in |
Minimum display |
0.1 mg |
Capacity |
320 g |
Repeatability (Standard deviation) |
0.15 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.0 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Not Included |
Sample (Concentration) Preparation |
Not Included |
mol Conversion Function |
Included |
Recipe Compounding |
Not Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Not Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.3 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) |
AP224X
External Calibration Weight Range for Span Calibration |
95 to 220.009 g |
---|---|
Calibration Weight |
Built-in |
Minimum display |
0.1 mg |
Capacity |
220 g |
Repeatability (Standard deviation) |
0.1 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.0 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Not Included |
Sample (Concentration) Preparation |
Not Included |
mol Conversion Function |
Included |
Recipe Compounding |
Not Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Not Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) |
AP124X
External Calibration Weight Range for Span Calibration |
45 to 120.009 g |
---|---|
Calibration Weight |
Built-in |
Minimum display |
0.1 mg |
Capacity |
120 g |
Repeatability (Standard deviation) |
0.1 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.0 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Not Included |
Sample (Concentration) Preparation |
Not Included |
mol Conversion Function |
Included |
Recipe Compounding |
Not Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Not Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) |
AP324W
External Calibration Weight Range for Span Calibration |
95 to 320.009 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.1 mg |
Capacity |
320 g |
Repeatability (Standard deviation) |
0.15 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.3 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP224W
External Calibration Weight Range for Span Calibration |
95 to 220.009 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.1 mg |
Capacity |
220 g |
Repeatability (Standard deviation) |
0.1 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP124W
External Calibration Weight Range for Span Calibration |
45 to 120.009 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.1 mg |
Capacity |
120 g |
Repeatability (Standard deviation) |
0.1 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 213(W) x 367(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg |
Minimum Weight *2 |
200 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP225WD
External Calibration Weight Range for Span Calibration |
95 to 220.009 g |
---|---|
Calibration Weight |
Built-in *1 |
Minimum display |
0.01 mg / 0.1 mg |
Capacity |
220 g / 102g |
Repeatability (Standard deviation) |
0.1 mg / 0.05 mg |
Inspection Support Function |
Included |
Display |
OEL display (dot matrix) |
Weight |
Approx. 7.9 kg |
Body Dimensions |
Approx. 213(W) x 411(D) x 345(H)mm |
Pan Size |
ø91 mm |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Ionizer |
Optional |
Clock-CAL |
Included |
HPLC Buffer Solution Preparation |
Included |
Sample (Concentration) Preparation |
Included |
mol Conversion Function |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. / 8 sec |
Response Time for race Measurements *3 |
2 sec. |
Linearity *2 |
±0.2 mg /±0.1 mg |
Minimum Weight *2 |
30 mg |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |
AP324W-AD
Linearity *2 |
±0.3 mg |
---|---|
Minimum Weight *2 |
200 mg |
Repeatability (Standard deviation) |
0.15 mg |
External Calibration Weight Range for Span Calibration |
95 to 320.009 g |
Calibration Weight |
Built-in *1 |
Minimum display |
0.1 mg |
Capacity |
320 g |
Response Time for race Measurements *3 |
2 sec. |
Sensitivity Stability Against Temperature Range |
±2 ppm/°C (10 to 30°C) |
Ionizer |
Included |
Operating Temperature/Humidity Range |
5 to 40°C 20 to 85% *5 |
Display |
OEL display (dot matrix) |
Pan Size |
ø91 mm |
Body Dimensions |
Approx. 215(W) x 367(D) x 346(H)mm |
Weight |
Approx. 7.0 kg |
Touchless Sensors |
Included |
Adjustable Internal Windbreak Plate |
Not Available |
Sample (Concentration) Preparation |
Included |
Automatic Doors |
Included |
Clock-CAL |
Included |
Inspection Support Function |
Included |
mol Conversion Function |
Included |
HPLC Buffer Solution Preparation |
Included |
Recipe Compounding |
Included |
USB Device (Type B) |
Included |
USB Host (Type A) |
Included |
Response Time *4 |
2 sec. |
Input/Output Terminal |
RS-232C (D-sub9P plug) USB host (Type A) USB device (Type B) Ionizer |